Please use this identifier to cite or link to this item: https://hdl.handle.net/1/1014
Title: An EPID-based system for gantry-resolved MLC quality assurance for VMAT
Authors: Zwan, Benjamin ;Barnes, M.P.;Fuangrod, T.;Stanton, C.J.;O'Connor, D.J.;Keall, P.J.;Greer, P.B.
Affliation: Central Coast Local Health District
Gosford Hospital
Issue Date: Sep-2016
Source: 17(5):348-365
Journal title: Journal of Applied Clinical Medical Physics
Department: Central Coast Cancer Centre
Abstract: Multileaf collimator (MLC) positions should be precisely and independently measured as a function of gantry angle as part of a comprehensive quality assurance (QA) program for volumetric-modulated arc therapy (VMAT). It is also ideal that such a QA program has the ability to relate MLC positional accuracy to patient-specific dosimetry in order to determine the clinical significance of any detected MLC errors. In this work we propose a method to verify individual MLC trajectories during VMAT deliveries for use as a routine linear accelerator QA tool. We also extend this method to reconstruct the 3D patient dose in the treatment planning system based on the measured MLC trajectories and the original DICOM plan file. The method relies on extracting MLC positions from EPID images acquired at 8.41 fps during clinical VMAT deliveries. A gantry angle is automatically tagged to each image in order to obtain the MLC trajectories as a function of gantry angle. This analysis was performed for six clinical VMAT plans acquired at monthly intervals for three months. The measured trajectories for each delivery were compared to the MLC positions from the DICOM plan file. The maximum mean error detected was 0.07 mm and a maximum root-mean-square error was 0.8 mm for any leaf of any delivery. The sensitivity of this system was characterized by introducing random and systematic MLC errors into the test plans. It was demonstrated that the system is capable of detecting random and systematic errors on the range of 1-2 mm and single leaf calibration errors of 0.5 mm. The methodology developed in the work has potential to be used for efficient routine linear accelerator MLC QA and pretreatment patient-specific QA and has the ability to relate measured MLC positional errors to 3D dosimetric errors within a patient volume. PACS number(s): 87.55.Qr.
URI: https://elibrary.cclhd.health.nsw.gov.au/cclhdjspui/handle/1/1014
DOI: 10.1120/jacmp.v17i5.6312
Pubmed: https://www.ncbi.nlm.nih.gov/pubmed/27685132
ISSN: 1526-9914
Publicaton type: Journal Article
Keywords: Cancer
Radiotherapy
Appears in Collections:Oncology / Cancer

Show full item record

Page view(s)

42
checked on May 7, 2024

Google ScholarTM

Check

Altmetric


Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.